www.vtiinstruments.com
VM2710A Command Dictionary
77
CALibration:CONVert:GAIN:VOLTage[:DC]?
Purpose
Reports the calibration gain value
Type
Query
Command Syntax
N/A
Command Parameters
N/A
*RST Value
N/A
Query Syntax
CALibration:CONVert:GAIN:VOLTage:DC? <range>
Query Parameters
<range> = 0.1 V | 1.0 V | 10.0 V | 100.0 V | 300.0 V
Query Response
See Table 3-1: Function and Range
Description
Reports the gain for a specified range in the dc voltage function. See the Scale column
in Table 3-1: Function and Range.
Calibration commands should only be executed by qualified
personnel. Changing these values incorrectly can cause the
instrument to perform improperly.
Examples
Command / Query
Response (Description)
CAL:CONV:GAIN:VOLT:DC? 10
(Queries and reports the gain for the 10 V
range)
Related Commands
CALibration:CONVert:GAIN?
CALibration:CONVert:GAIN:CURRent:AC?
CALibration:CONVert:GAIN:CURRent[:DC]?
CALibration:CONVert:GAIN:FRESistance?
CALibration:CONVert:GAIN:RESistance?
CALibration:CONVert:GAIN:VOLTage:AC?