MEASure:VOLTage[:DC]?
Purpose
Selects the dc voltage function
Type
Setting
Command Syntax
MEASure:VOLTage[:DC] [<expected value>[<resolution>]]
Command Parameters
<expected value> = 0 V through ±300 V | AUTO | DEF | MIN | MAX
<resolution>
= 1e-6 | 1e-5 | 1e-4 | 1e-3 | 1e-2 | DEF | MIN | MAX
Refer to the
DC Voltage Specifications
*RST Value
DEFault - Autoranging / 1e-3
Query Syntax
N/A
Query Parameters
N/A
Query Response
N/A
Description
This command selects the dc voltage function. It also allows you to set parameters for
an expected voltage and resolution. When a specific voltage value is specified, the
multimeter will automatically selects the correct range. If autoranging is selected, the
only valid parameters for resolution are
MINimum
or
MAXimum
. To input a
resolution parameter, a voltage parameter must be specified first. Valid parameter
combinations are: no parameter, a voltage parameter only, or both a voltage parameter
and a resolution parameter. See the
DC Voltage Specifications
in Section 1 for allowed
range and resolution values. The input parameters are as follows:
Voltage
0 V - ±300 V
Specific expected voltage value
AUTOmatic
Autoranging
DEFault
Autoranging
MINimum
0 V
MAXimum
±300 V
[no value specified]
Autoranging
Resolution
1e-6 | 1e-5 | 1e-4 | 1e-3 | 1e-2
A value specified for resolution
DEFault
1e-3
MINimum
(Dependant on Range)
MAXimum
(Dependant on Range)
[no specified value]
1e-3 (Dependant on Range)
Examples
Command / Query
Response (Description)
CONF:VOLT:AC MAX,MAX
(Selects ±300 V at 1 mV resolution.)
CONF:VOLT:AC 12,1E-2
(Selects the 100 V range at 10 mV resolution.)
Related Commands
READ?