www.vtiinstruments.com
VM2710A Command Dictionary
129
[SENSe:]CURRent:NPLC
Purpose
Sets the integration time in power line cycles (PLCs)
Type
Setting
Command Syntax
[SENSe:]CURRent:NPLC <nplc>
Command Parameters
<nplc> = 0.03 | 0.1 | 1 | 10 | 100 | DEF | MIN | MAX
*RST Value
1
Query Syntax
[SENSe:]CURRent:NPLC? [<MIN | MAX>]
Query Parameters
<MIN> = Causes the query to report the minimum value (0.00)
<MAX> = Causes the query to report the maximum value (100)
Query Response
0.03 | 0.1 | 1 | 10 | 100
Description
This command sets the integration time, for current measurements, in power line cycles
(PLCs). Values are rounded up to the nearest number of PLCs as shown in the table
below. MINimum sets the value to 0.03 PLC; MAXimum sets the value to 100 PLC.
The greater number of PLCs, the greater the normal mode rejection and the lower the
reading rate. Use the CALibration:LFRequency command to select the line frequency
reference for the multimeter's A/D converter.
Power Line Cycles (PLCs)
Frequency
0.03 PLC
50 Hz | 60 Hz
0.1 PLC
50 Hz | 60 Hz
1 PLC
50 Hz | 60 Hz
10 PLC
50 Hz | 60 Hz
100 PLC
50 Hz | 60 Hz
Setting integration time in power line cycles also sets the aperture time and the
resolution. This command can override the previous values set by CURRent:NPLC or
CURRent:RESolution. The last command executed has priority.
Examples
Command / Query
Response (Description)
CURR:NPLC 10
CURR:NPLC?
10
Related Commands
CALibration:LFRequency
[SENSe:]CURRent:APERture
[SENSe:]CURRent:RESolution