VM2710A Command Dictionary
113
MEASure:VOLTage:AC?
Purpose
Selects the ac-coupled rms voltage function
Type
Setting
Command Syntax
MEASure:VOLTage:AC [<expected value>[, <resolution>]]
Command Parameters
<expected value> = 0 V through ±300 V | AUTO | DEF | MIN | MAX
<resolution>
= 1e-6 | 1e-5 | 1e-4 | 1e-3 | 1e-2 | DEF | MIN | MAX
Refer to the
AC Voltage Specificaiotns
*RST Value
DEFault - Autoranging / 1e-3
Query Syntax
N/A
Query Parameters
N/A
Query Response
N/A
Description
This command selects the ac-coupled rms voltage function. It also allows you to set
parameters for an expected voltage and resolution. When a specific voltage value is
specified, the multimeter will automatically selects the correct range. If autoranging is
selected, the only valid parameters for resolution are
MINimum
or
MAXimum
. To
input a resolution parameter, a voltage parameter must be specified first. Valid
parameter combinations are: no parameter, a voltage parameter only, or both a voltage
parameter and a resolution parameter. See the
AC Voltage Specificaiotns
for range and
resolution values allowed. The input parameters are as follows:
Voltage
0 V - ±300 V
Specific expected voltage value
AUTOmatic
Autoranging
DEFault
Autoranging
MINimum
0 V
MAXimum
±300 V
[no value specified]
Autoranging
Resolution
1e-6 | 1e-5 | 1e-4 | 1e-3 | 1e-2
A value specified for resolution
DEFault
1e-3
MINimum
(Dependant on Range)
MAXimum
(Dependant on Range)
[no specified value]
1e-3 (Dependant on Range)
Examples
Command / Query
Response (Description)
MEAS:VOLT:AC MAX,MAX
(Selects ±300 V at 1 mV resolution.)
MEAS:VOLT:AC 12,1E-2
(Selects the 100 V range at 10 mV resolution.)
Related Commands
READ?