VTI Instruments Corporation
18
VM2710A Introduction
AC
C
URRENT
S
PECIFICATIONS
M
EASUREMENT
M
ETHOD
Direct coupled to fuse and shunt, ac-coupled true rms measurement
C
URRENT
R
ANGES
10 mA, 100 mA, and 1.0 A rms
O
VER
-R
ANGE
20%
S
HUNT
R
ESISTORS
1.0 A Range
10 mA & 100 mA Range
0.1 Ω
8.1 Ω
B
URDEN
V
OLTAGE
10 mA
100 mA
1.0 A
≤ 100 mV
≤ 1.0 V
≤ 120 mV
O
VER
-
CURRENT
P
ROTECTION
2.0 A fuse
R
ESOLUTION
Range
100 PLC
(6.5 Digits)
10 PLC
(6.5 Digits)
1 PLC
(5.5 Digits)
0.1 PLC
(5.5 Digits)
0.03 PLC
(4.5 Digits)
10 mA
100 nA
100 nA
1
µ
A
1
µ
A
1
µ
A
100 mA
1
µ
A
1
µ
A
1
µ
A
10
µ
A
10
µ
A
1.0
A
10
µ
A
10
µ
A
10
µ
A
100
µ
A
100
µ
A
24
HR
A
CCURACY
1
Range
10 mA Range
100 mA Range
1.0 A Range
20 Hz – 30 Hz
1.00 + 0.20
1.00 + 0.04
1.00 + 0.20
30 Hz – 50 Hz
0.50 + 0.20
0.50 + 0.04
0.50 + 0.20
50 Hz – 1k Hz
0.20 + 0.20
0.20 + 0.04
0.20 + 0.20
1k Hz – 5kHz
0.17 + 0.20
0.17 + 0.20
0.17 + 0.20
5kHz – 10kHz
0.55 + 0.30
0.55 + 0.30
3.0 + 0.30
90
DAY
A
CCURACY
2
Range
10 mA Range
100 mA Range
1.0 A Range
20 Hz – 30 Hz
1.10 + 0.20
1.10 + 0.05
1.10 + 0.20
30 Hz – 50 Hz
0.60 + 0.20
0.60 + 0.05
0.60 + 0.20
50 Hz – 1kHz
0.30 + 0.20
0.30 + 0.05
0.30 + 0.20
1kHz – 5kHz
0.27 + 0.20
0.27 + 0.20
0.27 + 0.20
5kHz – 10kHz
0.65 + 0.30
0.65 + 0.30
3.1 + 0.30
1
YR
A
CCURACY
2
Range
10 mA Range
100 mA Range
1.0 A Range
20 Hz – 30 Hz
1.12 + 0.20
1.12 + 0.06
1.12 + 0.20
30 Hz – 50 Hz
0.62 + 0.20
0.62 + 0.06
0.62 + 0.20
50 Hz – 1kHz
0.32 + 0.20
0.32 + 0.06
0.32 + 0.20
1kHz – 5kHz
0.29 + 0.20
0.29 + 0.20
0.29 + 0.20
5kHz – 10kHz
0.67 + 0.30
0.67 + 0.30
3.12 + 0.30
T
EMPERATURE
C
OEFFICIENT
:
0
°C
-
18
°C
AND
28
°C
-
55
°C;
(%R
EADING
+
%R
ANGE
)
/
°C
20 Hz – 30 Hz
30 Hz – 50 Hz
50 Hz – 10kHz
0.020 + 0.001
0.010 + 0.001
0.006 + 0.001