VTI Instruments Corporation
66
VM2710A Command Dictionary
CALCulate:[FUNCtion:]LIMit:UPPer
Purpose
Sets the upper limit value
Type
Setting
Command Syntax
CALCulate:[FUNCtion:]LIMit:UPPer <limit>
Command Parameters
<limit> = number | MAXimum | MINimum | LAST
*RST Value
0
Query Syntax
CALCulate:[FUNCtion:]LIMit:UPPer?
Query Parameters
N/A
Query Response
Number
Description
This command sets the upper reading limit for the current function and range. If a
measurement is above this set limit, the limit high failure bit would be set in the Status
Questionable register:
Limit High Failure:
0x1000
Limit High Failure Pulse: 0x4000
Examples
Command / Query
Response (Description)
CALC:LIM:UPP 8.8
(Sets the upper limit to 8.8 V in the 10 V range)
CALC:LIM:UPP LAST
(Sets the upper limit to the value of the last
measurement)
Related Commands
CALCulate:FUNCtion
CALCulate:[FUNCtion:]LIMit:LOWer
CALCulate:[FUNCtion:]LIMit:UPPer:TTLTrg