Output function
0% value
100% value
Wre526
0.0 °C
2000.0 °C
Wre325
0.0 °C
2000.0 °C
Pt10
0.0 °C
500.0 °C
PT100
0.0 °C
400.0 °C
Cu50
0.0 °C
150.0 °C
Cu100
0.0 °C
150.0 °C
b) Automatic ramp function
The calibrator enables automatic output of the output values using the ramp function. It enables
to carry out linear tests very easily. Ranges are output continuously from 0% - 100% - 0%.
There are three modes available:
0% - 100% - 0%, smooth ramp, rise/fall time 40 s
0% - 100% - 0%, smooth ramp, rise/fall time 15 s
0% - 100% - 0%, step ramp with 25% increments, 5-second pause after each incre-
ment
To enable this function, briefly press the “
” button (15). Each keystroke switches to
the next ramp function.
Pressing and holding down the “
” button (15) (>1.5 s) disables the function.
c) Turning the display illumination on and off
Briefly pressing the power button (4) turns the display illumination on and off.