17
DMV-100 Rotor Creep and Speed Detector - Installation Manual
5. Attach the SMA plug from the sensor integral
cable to the SMA jack at the end of the
triaxial extension cable. Use two 8mm (or
[5/16”]
) flat wrenches to tighten the
connection firmly.
Recommended connection
torque: 1.7 Nm
[15 in-lb
]
Figure 11 : Tightening the SMA plug and SMA jack connection
6. Slide the heat-shrinkable tubing over the
connection and shrink the tubing with a heat
gun.
The heat-shrinkable tubing must
be placed over the connection to
prevent accidental short-circuits
between connectors and metallic
parts such as flexible conduit, stator
frame, etc., and to secure the con-
nector.
Figure 12 : Heating the heat-shrinkable tubing
7. Pull the connection inside the tubing,
just enough to straighten the cable.
If the sensor integral cable is
equipped with a ground wire, also pull
this wire through the protective
tubing.
Figure 13 : Pulling the SMA connection inside the protective tubing