Chapter 16
Maintenance and Troubleshooting
Troubleshooting
156
User Manual
76100000200 Rev. 001
4
Enter the password
02468753
and confirm clicking on
Enter
A window displays during deletion and once the deletion is
completed, a message inform you.
5
Press any key to continue.
Touchscreen
calibration
If you meet problems when using the touchscreen (example: pressing on
an icon does not work correctly...) a calibration can be performed on the
6000A V2 Platform touchscreen. To calibrate the touchscreen, proceed
as follows:
1
In the
Home
page, press
System Settings menu key.
2
In the
System Settings
page, press menu key
Touch Screen
.
A blue slightly smaller screen appears, displaying a little target on
the left hand corner.
3
Click on this target (preferably with the stylus for touchscreen).
4
A new target appears then and again, for a total of 4 times, in order
to click on all corners of the screen.
If all the targets are not correctly touched, the touchscreen cannot be
used. Otherwise, the touchscreen may be used directly.
Changing the
battery
If you meet problems during the Platform functioning, or if the battery
does not charge anymore when plugged, this may require the battery to
be replaced
Removing the
damaged battery
To access the battery of the 6000A V2 Platform, proceed as follows
1
Switch off the 6000A V2 Platform and disconnect the mains supply.
2
Turn the instrument face down on the work surface.
Fig. 102
Touchscreen calibration
Target
Summary of Contents for T-BERD 6000A V2
Page 2: ......
Page 4: ......
Page 16: ...Table of Contents xvi User Manual 76100000200 Rev 001 ...
Page 68: ...Chapter 6 Scope File menu 48 User Manual 76100000200 Rev 001 ...
Page 78: ...Chapter 7 Applications File Explorer 58 User Manual 76100000200 Rev 001 ...
Page 114: ...Chapter 10 File management Storage media 94 User Manual 76100000200 Rev 001 ...
Page 187: ......