OMMANDS
9 R
EMOTE
C
ONTROL
Extended commands for the Probe application/PCM
application
:STOR:GRP:NAMES?
This setup contains the names of all available storage groups. The
“default” group is always available.
:STOR:NAMES?
This setup contains the fiber IDs and internal IDs belonging to the
group specified by :STOR:GRP:NAME.
:STOR:NOTE
<STRING>
A user specific note which is stored with the measurement results.
:STOR:NOTE?
Returns the stored user note.
Response Type:
<STRING>
See :STOR:NOTE command for more information.
:STOR:WR:BUSY?
Returns TRUE if a storage command is active. After saving the
measurement the value FALSE will be returned.
Response type:
<BOOLEAN>
Command string
Parameter type / Response type / Unit / Info
Command string
Parameter type / Response type / Unit / Info
:APP:FACT:DEF
Sets all parameters of the application to their default values.
:MICR:CAL:INFO?
Returns the list of calibrations (“tips”) of the connected Digital Probe.
Example of a list: (sf = supported fibers)
PCM: 2 active calibrations: [0], “Standard Tips (with BAP1)”, sf=1;[1],
“Ribbon Tips”, sf=2
:MICR:MODL:NAME?
Returns the model of the connected Digital Probe. Usually: “FBP-
P5000i”
:MICR:SER:NUMB?
Returns the serial number of the connected Digital Probe. For example
“B1712-8740-399”.
:MICR:SW:VER?
Returns the firmware version of the connected Digital Probe. For
example “1.2.41.903”.
:STOR:ALL:DEL
Deletes all measurement entries and groups. Creates a new empty
“default” group.
:STOR:GRP:NAME is set to default.
:STOR:DEL
<STRING>,
<STRING>,
<STRING>
Deletes a dedicated measurement result specified by group (string1),
fiber identifier (string2) and internal identifier (string3).