7. MAINTENANCE
Any maintenance work not detailed in this handbook should be carried
out by an approved supplier or distributor. If further information is
required on any aspect of maintenance please contact Customer
Service.
Identification of Consumables
There are six types of unique replacement consumables designed for
use in the
MEDICA-D 7/15 US
units and these are illustrated with the
following part numbers:
LC140
Pre-treatment cartridge
LC141
Ion-exchange cartridge
LC105
UV lamp
LC125
0.2µm capsule filter
LC136
Composite vent filter (for DV)
LC154
Degas module
To protect the inlet solenoid valve, RO boost pump (when fitted) and
recirc/delivery pump from possible debris in the water, the unit
incorporates two strainers.
WARNING!
ALWAYS
CHECK
THAT
THE
MAINS
ELECTRICAL POWER AND FEED WATER
SUPPLIES ARE SWITCHED OFF BEFORE
ATTEMPTING TO CHANGE THE
MEDICA-D
7/15
CONSUMABLES.
Frequency of Consumable
Replacement
The following frequency of consumable replacement is recommended
as a guide assuming typical usage.*
Pre-treatment
- LC140
maximum 6 months
Ion exchange
- LC141
maximum 6 months
UV lamp
- LC105
maximum 12 months
0.2µm capsule filter
- LC125
maximum 6 months
Composite vent filter - LC136
maximum 6 months
Reverse Osmosis
- LC143
every 2 - 3 years
(not an operator
replacement item)
Degas module
- LC154
every 2 - 3 years
(not an operator
replacement item)
*
These frequencies are only estimates and replacement will
depend on the application and feed water quality.
CAUTION!
Ensure that the display and replacement
timer settings are reset after replacing
consumables. (Refer to Section 5.4).
Pre-treatment
cartridge
Ion-exchange
cartridge
UV lamp
0.2µm capsule filter
MEDICA-D 7/15 US
Replacement Consumables
LC154