MEDICA-D 7/15 US
Operator Manual
ELGA
Page 30
MEDICA-D 7/15 US MANU40550 Version 2
– 05/17
Step 3 - Replacing the 0.2µm Capsule Filter
1. UNPACK new cartridge and remove the two
protective transit plugs sealing the inlet and outlet
connections.
2. REFIT the inlet tubing into the bottom of the filter
by pushing the tubing into the connectors until
locked and held by the retaining collet.
3. REFIT the outlet tubing.
4. SECURE the new filter into its retaining clips
ensuring the cartridge is the correct way up.
5. REFIT the pre-treatment cartridge by pushing it
back into its retaining clips.
6. RESET 0.2µm capsule filter replacement timer.
(See Section 5.5
– Setting Up Display and
Replacement Timer).
7. Follow steps 16 - 20 of
Section 5.6 Initial Start
Up
.
7.5 Cleaning the Inlet Strainer
The inlet strainer should be checked and cleaned every six months to
ensure that the strainer does not become clogged.
Step 1 - Remove the Inlet Strainer
1. SWITCH OFF electrical supply.
2. OPEN front door.
3. ISOLATE inlet water supply.
4. REMOVE the pre-treatment cartridge from its
clips and set aside to gain access to the inlet
strainer.
5. DEPRESS collars on both sides of strainer and
disconnect tubing.
6. REMOVE the inlet strainer from its position.
Step 2 - Dismantle the Inlet Strainer
1. HOLD inlet strainer over a sink or receptacle.
2. UNSCREW inlet strainer.
3. REMOVE mesh filter.
4. CHECK mesh filter for signs of wear or damage,
replace or clean as necessary.
Step 3 - Reassemble the Inlet Strainer
1. INSERT mesh filter into strainer, ENSURE it is
facing the correct direction.
2. TIGHTEN up the inlet strainer.
Step 4 - Replace the Inlet Strainer
1. REPOSITION the inlet strainer.
2. REFIT tubes to inlet strainer, ENSURE it is facing
the correct direction.
3. REPOSITION the pre-treatment cartridge into its
support clips.
4. RE-ESTABLISH inlet water supply.
5. TURN on power.
Inlet Strainer