17. REMOVE the by-pass block (Store in the front
door). INSERT ion-exchange cartridge LC141
(See Section 7.2 Step 3, 1-5). CLOSE front door.
OPEN the sanitization isolating valve (V
2
).
18. PRESS the PROCESS button to flush the LC141
ion-exchange cartridge.
19. Slowly open the upper bleed nipple on the 0.2µm
capsule filter until water, free of air bubbles, flows
out - use an absorbent cloth to prevent splashing
of internal components.
20. After 20 minutes PRESS the PROCESS button to
stop flow and close the sanitization isolating valve
(V2). PRESS the PROCESS button to start the
water purification. CHECK unit for leaks.
21. Allow the system to operate until the water purity
indicator displays a value within acceptable
operating values.
22. OPEN analyser isolating valve (V1).
23. The unit is now ready for use; it will run
automatically, fill the treated water docking
vessel/reservoir and recirculate intermittently.
24. When the analyzer calls for water (i.e. its inlet
valve is open) the system will operate and provide
water.
Bleed
nipple
0.2µm
capsule filter
Bleed Nipple