7.2 Replacing LC141 Ion-Exchange Cartridge Pack
The ion-exchange cartridge pack should be replaced in the following
circumstances:
The water purity alarm monitor indicates that water purity has
fallen below limits and the pack probably requires changing.
If the system is being re-commissioned or sanitized after an
extended period in which it was not used.
When indicated by the change reminder.
Immediately before and after use of the emergency by-pass
loop.
Step 1 - Switch Unit Off
1. SWITCH the
MEDICA-D 7/15 US
off at the power
switch at the top left hand side of the unit.
2. RELIEVE any residual pressure from the system,
by waiting several minutes before proceeding.
3. Slowly OPEN the bleed nipple on the capsule
filter until water flow ceases
4. USE an absorbent cloth to soak up the water and
RE-TIGHTEN the bleed nipple.
WARNING!
ENSURE THE UNIT IS ISOLATED BEFORE
REMOVING THE ION-EXCHANGE
CARTRIDGE.
Step 2 - Remove Ion-Exchange Cartridge
1. OPEN the front door.
2. PUSH on cartridge top cap.
3. LIFT up cartridge.
4. SLIDE out cartridge.
5. DISCARD used ion-exchange cartridge.
Note:
The consumable is non-hazardous.
Dispose of as ordinary waste, observing
all local and national regulations.
Ion-exchange
cartridge
Removal and Fitting of Ion-Exchange
Cartridge
Bleed
nipple
0.2µm
capsule filter
0.2µm Capsule Filter & Bleed Nipple