MEDICA-D 7/15 US
Operator Manual
ELGA
Page 18
MEDICA-D 7/15 US MANU40550 Version 2
– 05/17
5.5 Setting Up Display and Replacement Timers
1. TURN the unit off at the power inlet module.
2. To enter the replacement timer set-up, PRESS
the left hand touch pad button and at the same
time TURN the power on. Graphics of the ion-
exchange cartridge, UV lamp, pre-treatment
cartridge and 0.2µm capsule filter are displayed
with hourly timer status.
CAUTION!
Before re-setting any of the cartridge
timers, ensure that the appropriate new
cartridges
have
been
installed
and
securely located correctly in the
MEDICA-D
7/15 US.
Step 1 - Ion-Exchange Cartridge Replacement Timer
Setting this screen will cause the ion-exchange cartridge timer
to reset to the preset value of 4380 hours (6 months).
1. Press
to jump to the next consumable or
to
initiate reset.
2. Press
to reset timer or press to abort reset.
3. Press
to jump to the next consumable.
Step 2 - UV Lamp Replacement Timer
Setting this screen will cause the UV lamp timer to reset to the
preset value of 8760 hours (1 year).
1. Press
to jump to the next consumable or
to
initiate reset.
2. Press
to reset timer or press to abort reset.
3. Press
to jump to the next consumable.
Step 3 - Pre-treatment Cartridge Replacement Timer
Setting this screen will cause the pre-treatment cartridge timer
to reset to the preset value of 4380 hours (6 months).
1. Press
to jump to the next consumable or
to
initiate reset.
2. Press
to reset timer or press to abort reset.
3. Press
to jump to the next consumable.
Pre-treatment Cartridge Timer
UV Lamp Replacement Timer
Ion-Exchange Cartridge Timer