MEDICA-D 7/15 US
Operator Manual
ELGA
Page 14
MEDICA-D 7/15 US MANU40550 Version 2
– 05/17
Pump feed
Recirculation
to docking
vessel/
reservoir
Outlet to
analyzer
Ejector return to
docking vessel/
reservoir
Right-Hand Side Connections
5.3 Connecting Up the
MEDICA-D 7/15 US
Once the
MEDICA-D 7/15 US
unit has been positioned either on a wall
or on a bench, it should be connected as follows:
Mains water inlet tube
Drain
Pump feed
Outlet to analyzer
Recirculation to docking vessel/reservoir
Ejector return to docking vessel/reservoir
Step 1 - Fitting Tubes
1. PUSH in collet on connector.
2. PULL out transit plug.
3. CUT a clean square end on an 8mm (5/16") O/D
semi rigid tube.
4. PUSH tube into connector.
CAUTION!
Do not restrict drain line.
CAUTION!
If the water supply is at a pressure greater
than 6bar (90psi) fit a pressure regulator
(LA512).
Pull plug
Connector
Collet
Blanking
transit plug
Depress
collet
Clean square
cut end
Tube 8mm OD
Simply push in tube to attach
Tube in secured position
Fitting Tubes
Feed Water Inlet
Connections
Blanking
plug
Drain
Feed
Water