Step 2
– Add sterilant into Sanitization Port
1. UNSCREW cap on sanitization port.
2. INSERT
ONE
EffetSan
™ tablet into the
sanitization port.
3. REFIT cap on sanitization port, hand tight.
Step 3 - Start the Sanitization Process
1. RESTORE the electrical supply.
2. PRESS the PROCESS button.
3. ALLOW the sanitization and rinsing to continue
for 2 hours.
4. PRESS the PROCESS button to stop the
process.
5. TURN OFF electrical supply.
Step 4 - Return to Normal Operation
1. Reconnect the pump feed to valve V3 and open
the isolating valve (V3).
2. TURN ON the electrical supply.
3. PRESS PROCESS button to return to normal
operation.
4. CHECK for leaks.
8.2.2
Sanitization Procedure for the unit and docking
vessel/reservoir
The unit and docking vessel/reservoir may be sanitized to destroy
bacteria within the pipework and filters of the unit. The complete
sanitization of the system is recommended to be performed if high
levels of bacterial contamination are found in the product water
feeding the analyzer. Should bacterial contamination of the product
water be suspected, samples should be taken by trained staff and
analyzed for bacterial counts. If the counts are abnormally high, ELGA
LabWater technical support or your local representative should be
contacted for advice and assistance.
It is recommended that only ELGA LabWater service personnel or
other fully trained staff should perform the procedure for sanitization of
the unit and docking vessel/reservoir.
flash
Sanitization Port
Insert
sanitant
.
Process On