Electrical
The units can be connected universally to any electrical supply
in the range of
100 - 240V and 50 - 60Hz.
The mains lead is
supplied with a molded plug on one end and a molded
connector to the unit on the other. The unit should be connected
to an earth.
Drain
A semi rigid flexible connection to a sink or suitable drain
capable of handling at least 1.5 l/min is required. The drain point
should have a gravity fall below the level of the unit and any
connections direct to drain should have an air-break device
fitted.
Feed Water
The feed water should be of good quality and comply with
specifications provided.
This should enter the unit via an
8mm
(5/16") O/D
semi rigid tube, and should be in the temperature
range 1 to 35°C.
CAUTION!
Operating temperatures outside the range
1 to 35°C will cause damage to the
MEDICA-D 7/15 US
unit.
For pressurized feeds, the minimum direct inlet pressure is
4.0bar (60psi) and maximum inlet pressure is
6bar (90psi).
Higher feed water pressures must be reduced using a pressure
regulator valve (Part No. LA512).
Mains power socket
Mains power lead
Fuse
ON/OFF
switch
Electrical Connections