33
6 Set up with the indicating and adjustment module
VEGAFLEX 83 • 4 … 20 mA/HART - two-wire
41834-EN-120726
by the user, particularly when setting the switching point on the level
switch.
Since the scaling is very comprehensive, the scaling of the level value
was divided into two menu items. In the menu item "
Level 1
" you
define the scaling size and the scaling unit for the level value on the
display, e.g. volume in l.
Since the scaling is very comprehensive, the scaling of the level value
was divided into two menu items.
In menu item "
Level 2
" you define the scaling format on the display
and the scaling of the level measured value for 0 % and 100 %.
Since the scaling is very comprehensive, the scaling of the interface
measured value was divided into two menu items. In the menu item
"
Interface 1
" you define the scaling size and the scaling unit for the
interface value on the display, e.g. volume in l.
You can accept the scaling of the level measurement also for the
interface measurement. If you select "
Yes
", the actual setting will be
displayed.
If you have selected "
No
", then you can enter the scaling for the
interface separately.
Setup/Scaling - Level 1
Setup/Scaling - Level 2
Setup/Scaling - Interface
1