8. Acquiring CEPH Images (Optional)
90
vatech A9 (Model name: PHT-30CSS) User Manual
5.
Adjust the instrument to the patient's sitting height using the Column UP/DOWN
on the Handle frame or the optional UP/DOWN switch.
After adjusting the height of the column, align the Ear Rods
to the patient.
6.
During the operation, correctly align the Ear Rods to the patient's ears, so his/her
head does not move.
7.
Carefully tilt the patient's head back and adjust so his/her Frankfurt Line is vertical
with the floor.
8.
Direct the patient to swallow first before closing his/her mouth and to remain in
his/her current position until image acquisition is completed.
9.
Click the
READY
button. The x-ray exposure has not started yet.
10.
Now go
to 8.4 X-ray Exposure
to start the exposure.