
12. Maintenance
136
vatech A9 (Model name: PHT-30CSS) User Manual
Item
Value
Voxel Size (0.20)
Standard
Gender/Age group
Man
X-ray Intensity
Hard
8.
Click the
Confirm
button.
9.
Click the
READY
button when enabled.
10.
Capture the PHANTOM image according to Console Software instructions.
11.
When image capturing is completed, save the image in EzDent-i.
12.4.2.2 Analyzing CT Uniformity Check Phantom
1.
Double-click the saved PHANTOM image in
EzDent-i
to run
Ez3D-i
.
2.
In the Axial pane, double click pane to maximize the Axial view.
4.
Make 25 mm lines from the center to UP/DOWN/LEFT/RIGHT directions as shown
below.
6.
Make the four boxes next to each 25 mm line as shown below.
Try to make each box as close to 25.0 mm2 in the area as possible.
Put the 4 ROIs at the end of each 25 mm line.
Summary of Contents for A9
Page 1: ......
Page 2: ...Notice 2 vatech A9 Model name PHT 30CSS User Manual...
Page 4: ...Notice 4 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 12: ...1 Introduction 12 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 18: ...2 General Information 18 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 124: ...10 Troubleshooting 124 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 140: ...12 Maintenance 140 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 170: ...15 Appendices 170 vatech A9 Model name PHT 30CSS User Manual...
Page 194: ......
Page 196: ......