9. Acquiring CBCT Images
110
vatech A9 (Model name: PHT-30CSS) User Manual
9.4.4 Patient Positioning – Third Molar (optional)
This feature is the only available option in some Asian
countries and Mexico.
1.
Insert the Third Molar Bite into the chin rest.
The Sanitary Vinyl Cover is for sole use only. It should be
replaced for each patient. Be sure to use the approved
vinyl cover.
Clean the Chinrest and the Bite with a non-alcohol-based,
non-corrosive cleaning solution and wipe with a dry towel
before the next patient.
The Third Molar mode is the same as the Normal mode in
the imaging and reconstruction method, however, to image
the patient's third molar, X-rays are exposed by rotating
the patient's head approximately 45 degrees. A special
byte is used for this.
2.
Guide the patient to a chair (sold separately) in the center of the equipment.
3.
Guide the patient to sit in the center of the equipment
If a chair with a backrest is used, make sure that the
patient's back is away from the backrest of the chair for
proper positioning.
Summary of Contents for A9
Page 1: ......
Page 2: ...Notice 2 vatech A9 Model name PHT 30CSS User Manual...
Page 4: ...Notice 4 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 12: ...1 Introduction 12 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 18: ...2 General Information 18 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 124: ...10 Troubleshooting 124 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 140: ...12 Maintenance 140 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 170: ...15 Appendices 170 vatech A9 Model name PHT 30CSS User Manual...
Page 194: ......
Page 196: ......