12. Maintenance
vatech A9 (Model name: PHT-30CSS) User Manual
137
Englis
h
7.
Measure the HOMOGENEITY by calculating the difference
between Max. average value and min. average value among the
4 ROIs and compare it with its standard.
CALCULATION
EVALUATION
Scope
Subtract the minimum average value
from the Maximum average value
among the 4 ROIs
The difference
should be less than
400 HU
61223.3.5
INTRODUCTION
8.
Record the mean CT Number value of each material.
10.
When finished, exit
EzDent-i.
12.4.3 High Contrast and Low Contrast Resolution Tests
Test Frequency
HIGH CONTRAST RESOLUTION: Initially and Yearly
LOW CONTRAST RESOLUTION: Initially and Yearly
Follow the High and Low Contrast Resolution Test procedure below to correctly image
and analyze the S&C Check Phantom. Test results must be documented and
maintained for at least one year.
12.4.3.1 Imaging S&C Check Phantom
1.
Remove Bite and Temple Supports from the Unit.
2.
Insert the PHANTOM JIG BOTTOM into the Chinrest and then put the PHANTOM
JIG TOP on the PHANTOM JIG BOTTOM.
3.
Make the PHANTOM JIG ASS’Y level by using BUBBLE LEVEL and three ALIGN
PINs.
4.
Put S&C CHECK PHANTOM on the PHANTOM JIG ASS’Y.
S&C CHECK PHANTOM
ID Number
Summary of Contents for A9
Page 1: ......
Page 2: ...Notice 2 vatech A9 Model name PHT 30CSS User Manual...
Page 4: ...Notice 4 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 12: ...1 Introduction 12 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 18: ...2 General Information 18 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 124: ...10 Troubleshooting 124 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 140: ...12 Maintenance 140 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 170: ...15 Appendices 170 vatech A9 Model name PHT 30CSS User Manual...
Page 194: ......
Page 196: ......