7. Acquiring PANO Images
68
vatech A9 (Model name: PHT-30CSS) User Manual
5.
Position the patient's back and cervical spine as straight as possible.
6.
Adjust the instrument to the patient's sitting height using the
Column UP/DOWN
on the Handle frame or the optional UP/DOWN switch while the patient holds the
handle.
7.
Close contact patient's chest to the handle frame.
8.
Guide the patient to press the base of the nose (acanthion point) against the
Chinrest and tilt the head forward about 5˚. At this point, make sure the patient's
jaw does not touch the equipment.
If the jaw touches the equipment, it is difficult to maintain
the proper position to get good images.
Be careful; the patient does not touch the equipment with
his/her jaw.
9.
Let the patient maintain the posture as follows:
Open the mouth.
Place the tongue on the roof of the mouth.
Close the eyes.
As shown in the picture, the support unit of the integrated
Chinrest should touch the patient's acanthion point.
Ask the patient to maintain his/her position until the
operation is completed.
Summary of Contents for A9
Page 1: ......
Page 2: ...Notice 2 vatech A9 Model name PHT 30CSS User Manual...
Page 4: ...Notice 4 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 12: ...1 Introduction 12 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 18: ...2 General Information 18 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 124: ...10 Troubleshooting 124 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 140: ...12 Maintenance 140 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 170: ...15 Appendices 170 vatech A9 Model name PHT 30CSS User Manual...
Page 194: ......
Page 196: ......