9. Acquiring CBCT Images
vatech A9 (Model name: PHT-30CSS) User Manual
113
Englis
h
9.4.5 Laser Beam Aligning – Third Molar (optional)
This feature is the only available option in some Asian
countries and Mexico.
Be careful not to shine the Laser Beam directly into the
person's eyes. Doing so may result in vision loss.
If the Laser Beam is not correctly positioned, there may
be distortion, causing the image to be enlarged or
reduced, or ghost shadows may occur and lower the
image quality. Be sure to align Laser Beam properly.
This is a sample illustration for reference only. Actual FOV
may vary from the image, as shown above.
2.
Align the center of the examinee's face according to the center alignment
guideline.
3.
Using a horizontal beam, correct the sagging or lifting of the patient's head.
9.4.6 Finishing Patient Positioning
1.
After checking the positions of the patient and the Laser Beam, prevent the
patient's head from moving.
2.
Click the
READY
button. X-ray exposure has not started yet.
3.
Now go to
9.5 X-ray Exposure
to start the exposure.
Frankfurt Plane
Summary of Contents for A9
Page 1: ......
Page 2: ...Notice 2 vatech A9 Model name PHT 30CSS User Manual...
Page 4: ...Notice 4 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 12: ...1 Introduction 12 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 18: ...2 General Information 18 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 124: ...10 Troubleshooting 124 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 140: ...12 Maintenance 140 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 170: ...15 Appendices 170 vatech A9 Model name PHT 30CSS User Manual...
Page 194: ......
Page 196: ......