12. Maintenance
vatech A9 (Model name: PHT-30CSS) User Manual
129
Englis
h
12.3 QA Test
This section is intended to give information about PHT-30CSS PHANTOM KIT for 3D
IMAGE QUALITY INSPECTION and CALIBRATION. It is recommended that you
thoroughly familiarize yourself with this guide to perform the regular QA test effectively
by using the PHANTOM KIT. QA tests should be done annually at the frequency
specified by the manufacturer or state regulations in which the X-ray system is being
used.
12.3.1 Phantom Kit Contents
CT NUMBER CHECK PHANTOM x 1
UNIFORMITY CHECK PHANTOM x1
S&C CHECK PHANTOM x1
PHANTOM JIG A’SSY x 1
PHANTOM KIT User Manual x 1
12.3.2 Specifications of Phantom Kit Contents
CT Number Check Phantom
Manufacturer
-
VATECH Co., Ltd.
Intended Use
-
CT Number Inspection
-
CT Number Calibration
Complies with
- IEC 61223-2-6
- IEC 61223-3-5
Uniformity Check Phantom
Manufacturer
-
VATECH Co., Ltd.
Intended Use
-
CT Image Homogeneity Inspection
-
CT Image Noise Inspection
Complies with
- IEC 61223-2-6
- IEC 61223-3-5
Summary of Contents for A9
Page 1: ......
Page 2: ...Notice 2 vatech A9 Model name PHT 30CSS User Manual...
Page 4: ...Notice 4 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 12: ...1 Introduction 12 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 18: ...2 General Information 18 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 124: ...10 Troubleshooting 124 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 140: ...12 Maintenance 140 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 170: ...15 Appendices 170 vatech A9 Model name PHT 30CSS User Manual...
Page 194: ......
Page 196: ......