12. Maintenance
128
vatech A9 (Model name: PHT-30CSS) User Manual
12.2 Maintenance Task Checklist
Tasks
Period
Before the operation, ensure that the equipment is clean and ready for
use. Make sure that all parts that come in direct contact with the patient
have been cleaned thoroughly.
Daily
After using the equipment, make sure that the
Main Power Switch
has
been turned off.
Daily
Ensure that the equipment is firmly plugged into a dedicated power
source.
Daily
Ensure that the plug and the power cord are not heated abnormally.
Daily
Confirm that the LED indicator turns yellow when
the Exposure Switch
is pressed. Ensure that the LED indicator remains yellow for the entire
duration of the exposure.
Daily
Ensure that the power cable is not kinked, broken, exposed, and free of
all other defects.
Daily
Confirm that activating the
Emergency Stop Switch
ceases the unit’s
operation. Pressing the
Emergency Stop Switch
should stop all
movement of the equipment and X-ray emission.
Weekly
Ensure that all visible labels are intact and legible.
Weekly
Check for damages to the
Exposure Switch
cable.
Monthly
Confirm that the audio message is audible throughout the exposure.
Monthly
Summary of Contents for A9
Page 1: ......
Page 2: ...Notice 2 vatech A9 Model name PHT 30CSS User Manual...
Page 4: ...Notice 4 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 12: ...1 Introduction 12 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 18: ...2 General Information 18 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 124: ...10 Troubleshooting 124 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 140: ...12 Maintenance 140 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 170: ...15 Appendices 170 vatech A9 Model name PHT 30CSS User Manual...
Page 194: ......
Page 196: ......