7. Acquiring PANO Images
72
vatech A9 (Model name: PHT-30CSS) User Manual
TMJ CLOSE Mode (PA) and Sinus Mode (LAT)
Patient Positioning
1.
Remove the
Normal Chinrest
and insert the
Special Chinrest
into the
equipment.
2.
Insert the
Special Bite B
into the
Special Chinrest
Clean the Chinrest and the Bite with a non-alcohol-based,
non-corrosive cleaning solution and wipe with a dry towel
before the next patient.
3.
Guide the patient to a chair (sold separately) in the center of the equipment.
4.
Guide the patient to sit in the center of the equipment
If a chair with a backrest is used, make sure that the
patient's back is away from the backrest of the chair for
proper positioning.
5.
Position the patient's back and cervical spine as straight as possible.
6.
Adjust the instrument to the patient's sitting height using the
Column UP/DOWN
on the Handle frame or the optional UP/DOWN switch while the patient holds the
handle.
7.
Close contact patient's chest to the handle frame.
8.
Guide the patient to sit in the center of the equipment and maintain the position
described below.
Hold the handle firmly.
Make sure the patient's chest is in contact with the equipment.
Keep both shoulders parallel.
Straighten the patient's cervical spine and sit still.
Special Bite B
Special Chinrest
Summary of Contents for A9
Page 1: ......
Page 2: ...Notice 2 vatech A9 Model name PHT 30CSS User Manual...
Page 4: ...Notice 4 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 12: ...1 Introduction 12 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 18: ...2 General Information 18 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 124: ...10 Troubleshooting 124 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 140: ...12 Maintenance 140 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Page 170: ...15 Appendices 170 vatech A9 Model name PHT 30CSS User Manual...
Page 194: ......
Page 196: ......