15. Appendices
172
vatech A9 (Model name: PHT-30CSS) User Manual
15.2.3.2 CBCT Mode Results
Test Condition
Tested Mode
CBCT
Distance from focal point [m]
1~2
Applied Tube Voltage Peak [kVp]
99
Applied Tube Current [mA]
12
Applied Exposure time [sec]
15.5
Mode
Direction [°]
CBCT FOV 8x8 [mR]
1 m
(3.3 ft)
1.5 m
(4.9 ft)
2 m
(6.6 ft)
0
Occiput
0.517
0.307
0.166
45
0.502
0.229
0.133
90
Left ear
0.527
0.242
0.134
135
0.466
0.253
0.147
180
Nose
0.113
0.066
0.042
225
0.502
0.254
0.154
270
Right ear
0.500
0.239
0.135
315
0.483
0.276
0.158