background image

 

beam hits the center of mirror M3 then use M3 so the beam hits the center of the 
sample.  Some of the mirrors are more complicated to adjust (although these can 

often be left alone).  For help with alignment, talk to one of the GLAs and read 
the alignment instructions, which can be found under Help. 

Small samples 

Most typical lab samples are smaller in size than what the instrument was designed to 

measure.  For all three types of measurements (T, C, and R), it is best to under fill the 
sample area.  

Spot size

 - There are two ways of 

reducing the beam spot size: 

1) 

Slit width:

 The simplest way is 

to reduce beam size is by the slit 
width.  Click on the 

Setup

 the 

options

 tab.  Under slit height, 

switch to Reduced.  It takes about a 
minute for the slit to change. 
2) 

Small spot kit:

 There are also a 

series of lenses that can be put in 
the optical path to focus the beam 

into a smaller spot on the sample. 
Again there are three lenses with 
different focal lengths and marked 

T, C, and R for the three positions of 
the integrating sphere.  Also, the 

large mirror (M3) is switched out 
for a one with a different curvature; 
this mirror is always kept in the T 

position, regardless of the type of 
measurement being made.   

3) Aperture kit for reflection 

measurements 

For normal incidence reflection measurements, the open circle in the rear of the 

integrating sphere is several inches in diameter.  There are a series of apertures, 
which have the white Spectralon material (same roughened Teflon that is inside the 
integrating sphere), that can slide onto the slot on the back of the integrating sphere.  

The spring-loaded mount that holds the sample against the integrating sphere first 
needs to be taken off.  Slide on the appropriate aperture and then either tape your 

sample onto the back of the aperture, or put back on the spring-loaded sample 
mount. 
If the beam spot overfills the aperture, you can correct for this “stray light” by 

performing a Zero/baseline Correction.  Normally, the 0%TBaseline scan is run by 
blocking the beam before it enters the integrating sphere, which accounts for 

Figure 4 Photo of diffuse reflectance accessory with 
beam path. 

Summary of Contents for Cary 5000

Page 1: ...n on computer 2 Check lens and mirror are correct for your measurement a Install Small Sample Kit mirror M3 SSK or regular M3 mirror if needed b If using SSK install correct lens L2 3 Make sure cover...

Page 2: ...e Turn on the instrument 0 I switch on the lower left side of the front of the instrument In the Scan program watch the lower left corner to see the status of the instrument When the traffic light is...

Page 3: ...mounts in front of the sample see Figure 4 There are three lenses that can be used again for the three positons marked as T C or R Each time the sample configuration is changed the optics need to be...

Page 4: ...T C and R for the three positions of the integrating sphere Also the large mirror M3 is switched out for a one with a different curvature this mirror is always kept in the T position regardless of the...

Page 5: ...on located in the upper left hand corner defines the scan parameters The following useful tabs can be found under Setup Cary tab Here you can set the units of the x and y axes nm cm 1 etc and A T R et...

Page 6: ...be applied to a sample where the abscissa range of the sample exceeds that of the baseline If you change your abscissa range to one outside that of the baseline the Cary will display an error message...

Page 7: ...x at the end of a collect where you can enter the Batch file name for your data Note In the Scan application Prompt at Start and Prompt at End save information as a DSW file unless there are multi or...

Page 8: ...multiple graphs in the Graphics area You can double click on the Graphics area to toggle the display through graph and report to graph only The graph functions can also be accessed by clicking one of...

Page 9: ...ich Text Format file How to Export Collected Data If you want to export data from a Cary WinUV application into a third party application then save your data as an ASCII CSV file To do this 1 Click Fi...

Page 10: ...eck box is not checked 8 In the File name field type in the file name for your new batch file 9 Click Save to create the new batch file The current method will be stored with the batch files All the d...

Page 11: ...SBW will change automatically to maintain the required signal level If you are collecting data in double beam mode across the UV Vis and NIR regions with Fixed SBW set as the Measurement Mode then you...

Page 12: ...nes for another Scan run you simply open the file using either the File Open Data command or click the Baseline button in the Baseline tab of the Setup dialog box Note If you decide that you do not wa...

Page 13: ...gle bounce relative SRA accessories should not use this correction Select a Baseline Zero correction instead This correction is performed with the raw transmission data and due to the double bounce on...

Page 14: ...owever sometimes you may notice that the text has turned to gray and is in italics This indicates that the Cary is currently idling outside the abscissa range of the baseline file but that the current...

Page 15: ...ferences and Graph Preferences options 1 Graph menu options Print XY point Select this to print the cursor s current X and Y coordinates in the report area This option is only available when right cli...

Page 16: ...to Arrange Graphs Select Auto Arrange Graphs to view and arrange all graphs simultaneously Copy Graph Select this to copy the Graphics area onto the Windows Clipboard Paste to Graph Select this to pas...

Page 17: ...in traces You can resize the Name of Trace and File name fields by selecting and dragging the splitter bar between the fields File Name This field displays the file name and path of the trace If the n...

Page 18: ...il the font is changed again Trace Style This field allows you to define the way that your trace will appear on screen Points Each individual data point collected by the Cary is shown on the graph Poi...

Page 19: ...e right side of each graph box A single click on a sample name or example line within the legend will make it the focused or red scan Information about the trace will be displayed below the X axis Dou...

Reviews: