background image

 

the VATH accessory, please talk to one of the GLAs for the instrument first.   

For the 

normal or large sample setup

 there are three positions for the large standard 

M3 mirror (see Figure 3 andFigure 4)that correspond to the three sample positions 

(Transmission, Center, and Reflection).  Each of the mirror positions will focus the light 
onto the position of the sample.  The three different mirror positions are marked T 

(transmittance slot), C (center mount), and R (rear reflectance slot) in Figure 3. Thus, 
the mirror needs to be moved to the right position depending on the placement of your 
sample.  The cable for the DRA can get in the way of moving this mirror.  

Turn off the 

Cary first, if you need to unplug the DRA to move the mirror.

  

For the small sample setup (see small sample discussion below) the M3 mirror is 

replaced by the small sample kit mirror (M3 SSK) in the T position.  Lens (L2) mounts in 
front of the sample (see Figure 4).  There are three lenses that can be used again for the 
three positons marked as T, C or R.  
Each time the sample configuration is changed, the optics need to be realigned, such 
that the sample and reference beams are not clipped at the entrance to the integrating 
sphere and that the sample beam is centered on the sample.  

1.

 

With the top door to the DRA closed, under the 

commands

 menu click on 

Go to

and then type in 550 nm. When you darken the room lights, you should see a 

green beam, which you can use to align the beam onto the sample (Alternatively, 
if you have trouble seeing the beam: Block the sample and reference beams from 
entering the integrating sphere. Under the 

commands

 menu go to 

align

 to 

switch the gratings to zero order to send white light into the DRA.  The white 
light should not enter the integrating sphere)   

2.

 

Most of the mirrors have two knobs, which you can rotate to adjust the vertical 

and horizontal positions of the beam.  Start with mirror M2 and adjust so the 

Figure 3 Diffuse Reflectance Accessory with beam path. 

Summary of Contents for Cary 5000

Page 1: ...n on computer 2 Check lens and mirror are correct for your measurement a Install Small Sample Kit mirror M3 SSK or regular M3 mirror if needed b If using SSK install correct lens L2 3 Make sure cover...

Page 2: ...e Turn on the instrument 0 I switch on the lower left side of the front of the instrument In the Scan program watch the lower left corner to see the status of the instrument When the traffic light is...

Page 3: ...mounts in front of the sample see Figure 4 There are three lenses that can be used again for the three positons marked as T C or R Each time the sample configuration is changed the optics need to be...

Page 4: ...T C and R for the three positions of the integrating sphere Also the large mirror M3 is switched out for a one with a different curvature this mirror is always kept in the T position regardless of the...

Page 5: ...on located in the upper left hand corner defines the scan parameters The following useful tabs can be found under Setup Cary tab Here you can set the units of the x and y axes nm cm 1 etc and A T R et...

Page 6: ...be applied to a sample where the abscissa range of the sample exceeds that of the baseline If you change your abscissa range to one outside that of the baseline the Cary will display an error message...

Page 7: ...x at the end of a collect where you can enter the Batch file name for your data Note In the Scan application Prompt at Start and Prompt at End save information as a DSW file unless there are multi or...

Page 8: ...multiple graphs in the Graphics area You can double click on the Graphics area to toggle the display through graph and report to graph only The graph functions can also be accessed by clicking one of...

Page 9: ...ich Text Format file How to Export Collected Data If you want to export data from a Cary WinUV application into a third party application then save your data as an ASCII CSV file To do this 1 Click Fi...

Page 10: ...eck box is not checked 8 In the File name field type in the file name for your new batch file 9 Click Save to create the new batch file The current method will be stored with the batch files All the d...

Page 11: ...SBW will change automatically to maintain the required signal level If you are collecting data in double beam mode across the UV Vis and NIR regions with Fixed SBW set as the Measurement Mode then you...

Page 12: ...nes for another Scan run you simply open the file using either the File Open Data command or click the Baseline button in the Baseline tab of the Setup dialog box Note If you decide that you do not wa...

Page 13: ...gle bounce relative SRA accessories should not use this correction Select a Baseline Zero correction instead This correction is performed with the raw transmission data and due to the double bounce on...

Page 14: ...owever sometimes you may notice that the text has turned to gray and is in italics This indicates that the Cary is currently idling outside the abscissa range of the baseline file but that the current...

Page 15: ...ferences and Graph Preferences options 1 Graph menu options Print XY point Select this to print the cursor s current X and Y coordinates in the report area This option is only available when right cli...

Page 16: ...to Arrange Graphs Select Auto Arrange Graphs to view and arrange all graphs simultaneously Copy Graph Select this to copy the Graphics area onto the Windows Clipboard Paste to Graph Select this to pas...

Page 17: ...in traces You can resize the Name of Trace and File name fields by selecting and dragging the splitter bar between the fields File Name This field displays the file name and path of the trace If the n...

Page 18: ...il the font is changed again Trace Style This field allows you to define the way that your trace will appear on screen Points Each individual data point collected by the Cary is shown on the graph Poi...

Page 19: ...e right side of each graph box A single click on a sample name or example line within the legend will make it the focused or red scan Information about the trace will be displayed below the X axis Dou...

Reviews: