1
User guide for Cary 5000 absorption spectrometer
with external DRA 1800 attachment.
(last updated 03/17/2016)
This guide is for use of the Cary 5000 with the DRA only. For use without the DRA as a
standard transmission spectrometer see the User guide for Cary 5000 in absorption
mode.
Important warnings!
•
Do not unplug or plug in the external DRA attachment when the instrument is
on. This will ruin the detectors of the integrating sphere
•
Do not put white light into the DRA with the DRA electrically connected to the
Cary.
•
Minimize the amount of time that the integrating sphere is open. Long-term light
exposure hurt the detector over time.
•
Wear clean gloves while using the instrument
•
Do not use liquid or powder samples in the integrating sphere.
Quick Start for using DRA
1.
Turn on computer
2.
Check lens and mirror are correct for your measurement.
a.
Install Small Sample Kit mirror (M3 SSK) or regular M3 mirror if needed
b.
If using SSK install correct lens L2.
3.
Make sure cover for the DRA is fully closed and turn Cary on.
4.
Click on the Cary icon on the taskbar
5.
Wait 20 minutes for the lamps to warm up.
6.
Select
Setup
a.
set wavelength range, %T or %R
b.
Under Options tab: set slit width and height, SBW to 2 nm, Detector and
Grating change wavelengths
c.
Under Baseline tab: chose zero/baseline
d.
Under Storage tab: Set your filename
7.
Make sure that all
exits
of the integrating sphere are blocked with full reflectors.
8.
Baseline
and run 100% T scan, then block the sample beam at the entrance to the
integrating sphere for the 0% T scan. If using an aperture holder for reflection
measurements, please see the “Aperture kit for small samples” section on page 4.
9.
Mount your sample
10.
Run
Spectra
11.
When done remove sample and remount reflectors.
12.
Turn off Cary