03-914998-55:6
53 of 63
Figure 6.3
6.3 Methodology
NOTE: The 4000 GCMS system will arrive with a Test Files CD containing
factory test method and data files. These method files can be used to pass the
sensitivity tests associated with the installation. Retain the test files CD for your
records.
Ensure the MS temperatures are set for this acquisition:
•
Trap: 150 °C
•
Manifold: 40 °C
•
Transfer Line: 230 °C
In System Control, activate the internal CI method from the Test Files CD.