Section 3. Preliminary Tuning
Before preliminary tuning, be sure the MS has returned to Analysis
temperatures and the GC column flow rate is set to 1 mL/min.
3.1 System Monitoring
After bakeout, go to Startup/Shut Down (shown in Figure 2.1) and monitor the
following:
•
Heated Zones
: Operating Conditions should be close to Current Set
Points.
•
Vacuum System: Status should be Ready, Pump Spin Speed should be
100%, Current should be significantly less than 300 mAmps, dependent
on damping gas flow rate.
•
Pneumatics (external mode only): Damping Gas On, Flow Rate 1.0 - 4.0
mL/min (should match the set point).
•
Getter
Co
ntrol
: Heater On, Temperature > 300 °C
Next, go to Diagnostics, select the Ion Source, Multiplier/Dynode, and Ion Gauge
check boxes (refer to Figure 3.1) and monitor the following:
•
Vacuum System
: Power should be 8-12 Watts with no damping gas,
12-15 Watts with damping gas flow.
•
Ion Gauge
: Pressure should read back a µTorr value that will vary
dependent on column flow, damping gas flow, air/water levels, etc. This
is the manifold pressure.
•
Trap
: Multiplier voltage operating condition should be displayed, Dynode
should be approximately 10,000V.
•
Ion Source (
external mode
)
: operating conditions should be displayed,
check for no read backs equal to zero. (NOTE: un-checking the Ion
Source box should cause Emission to read back < 5 µAmps or there
may be a current leakage issue indicating a need to clean the filament
and lens areas.)
•
Roughing Line Pressure
: should be significantly <100 mTorr.
•
Deviation
: should be less than 3 µA after auto tuning the electron and ion
lenses.
Summary of Contents for 4000 GCMS
Page 4: ...03 914998 55 6 4 of 63 This page intentionally left blank...
Page 10: ...03 914998 55 6 10 of 63 This page intentionally left blank...
Page 14: ...03 914998 55 6 14 of 63 Click Finish to close the wizard Figure 2 3...
Page 17: ...03 914998 55 6 17 of 63 Figure 2 6...
Page 18: ...03 914998 55 6 18 of 63 Figure 2 7...
Page 25: ...03 914998 55 6 25 of 63 Figure 3 2...
Page 27: ...03 914998 55 6 27 of 63 Figure 3 4 Figure 3 5...
Page 30: ...03 914998 55 6 30 of 63 Figure 3 8...
Page 34: ...03 914998 55 6 34 of 63 This page intentionally left blank...
Page 40: ...03 914998 55 6 40 of 63 Figure 5 6...
Page 41: ...03 914998 55 6 41 of 63 Figure 5 7...
Page 42: ...03 914998 55 6 42 of 63 Figure 5 8...
Page 50: ...03 914998 55 6 50 of 63 This page intentionally left blank...