03-914998-55:6
16 of 63
Figure 2.5
2.2 Quick Check of Air / Water
It is important to monitor the progress of the manifold pump down before initiating
a bakeout. Once the Pump Spin Speed reaches 100%, check for the presence of
a massive water peak (18 and 19 m/z, unlikely to be resolved) and the absence
of a nitrogen peak (28 m/z).
Only turn the trap on for a short period of time
(<5 seconds) so the filaments and multiplier are not adversely affected.
NOTE: This check of air water may not give usable results until after auto tuning
the system.
Go to Manual Control and click the
Active Segment
tab. Enter a Mass Range of
15 to 45 m/z (refer to Figure 2.6, Full Scan Parameters tab). Also enter a Fixed
Ionization Time of 100 microseconds for internal mode or 5000 microseconds for
external mode (refer to Figure 2.7, Ionization Control tab). Click
Activate
Changes
. Alternatively, one can use segment 1 of
C:\VarianWS\4000Service\4000 MS (Int or Ext) Service.mth.
This is a quick check to confirm the absence of a m/z 28 signal. The m/z 28
signal should not be present due to the massive intensity of the 18/19 signal
before bakeout. Presence of a m/z 28 signal can indicate a leak or contaminated
gas.
Summary of Contents for 4000 GCMS
Page 4: ...03 914998 55 6 4 of 63 This page intentionally left blank...
Page 10: ...03 914998 55 6 10 of 63 This page intentionally left blank...
Page 14: ...03 914998 55 6 14 of 63 Click Finish to close the wizard Figure 2 3...
Page 17: ...03 914998 55 6 17 of 63 Figure 2 6...
Page 18: ...03 914998 55 6 18 of 63 Figure 2 7...
Page 25: ...03 914998 55 6 25 of 63 Figure 3 2...
Page 27: ...03 914998 55 6 27 of 63 Figure 3 4 Figure 3 5...
Page 30: ...03 914998 55 6 30 of 63 Figure 3 8...
Page 34: ...03 914998 55 6 34 of 63 This page intentionally left blank...
Page 40: ...03 914998 55 6 40 of 63 Figure 5 6...
Page 41: ...03 914998 55 6 41 of 63 Figure 5 7...
Page 42: ...03 914998 55 6 42 of 63 Figure 5 8...
Page 50: ...03 914998 55 6 50 of 63 This page intentionally left blank...