03-914998-55:6
24 of 63
Figure 3.1
3.2 RF Ramp Adjustment
Go to the Checks and Adjustments tab in Manual Control, highlight RF Ramp
Adjustment and click
Start
(refer to Figure 3.2). Adjust the RF tuning rod as
shown in Figure 3.3 to minimize the Base Amount value displayed in the upper
right corner of the spectral window. The graphic display can also be monitored
for minimizing the intensity of the high end. Click
Done
when finished.