90
5. TH E
VOLCANO MEDIC 2
H OT A IR G EN ER ATO R
5.4. Commissioning
Do not leave the device unat-
tended while it is being opera-
ted. Put the Hot Air Generator
on a flat and stable surface.
Check whether the mains voltage cor-
responds to the voltage rating stated
on the device. The Power Cord and
the plug must be in perfect condition.
Only connect the Hot Air Generator to
a properly installed socket.
The device is safely disconnected
from the mains only when the plug is
removed from the socket.
When the power plug is plugged in, all
sequences of the LED display will il-
luminate briefly before displaying the
firmware number. The mains opera
-
tion display which then appears on the
display shows that the Hot Air Genera-
tor is connected to the power supply.
5.5. Heating Up
To switch on the heater, press the
touch button labeled ‘HEAT’. The set/
actual temperature will light up, signal-
ing that the heater is switched on. The
upper, orange display shows the ac-
tual temperature in the Filling Chamber
with the air pump switched on, and the
lower, white display shows the desired
set temperature. The heating process
is completed when the set value is
equal to the actual value. This is also
indicated acoustically when the pump
is switched on briefly. Heating up can
take up to two minutes depending
upon the chosen temperature. After
the heating process is completed, the
temperature is constantly maintained
at the set value.
For technical reasons the tem-
perature in the Filling Chamber
cannot be measured during
vaporization. During operation the
temperature is measured at the heat
exchanger. The display indicates the
reference values for the attached Fill-
ing Chamber with the air pump
switched on.