122
Problem
Possible cause/remedy
No aerosol is formed
Check whether the set and actual temperatures are
the same (at least 180°C (356°F)).
Check whether the Filling Chamber and/or Valve
Balloon or Tube are correctly attached.
Check whether dronabinol has dropped onto the Drip
Pad or whether there is fresh, unused cannabis floss
in the Filling Chamber.
The Valve Balloon does
not fill up.
Check whether the pump is turned on.
Check whether the Filling Chamber and/or Valve
Balloon are correctly attached.
The device switches off
(automatic switch-off).
The automatic switch off has intervened. Switch the
heater or pump on again.
Plant particles in the Val-
ve Balloon or Tube (when
using cannabis flos).
Check whether the Screen in the Filling Chamber Cap
is clogged or is not inserted into the notch properly.
Check whether the Filling Chamber was cleaned
according to the Instructions for Use.
10. HOT A IR G EN ER ATO R
Summary of Contents for Volcano MEDIC 2
Page 2: ......
Page 69: ...69 de en fr es it nl...