Parameter
Value
Data bits
7
Stop bits
1
Flow control
None
After startup in the STOP mode, the probe outputs the software version and the command
prompt.
HMP155 1.00
>
In the RUN mode, a measurement output starts immediately after startup.
In the POLL mode, the probe does not output anything after startup. See
4.2.1 Installing driver for USB cable
Before using the USB cable, install the USB driver on your computer. Follow the
instructions provided with the USB cable.
1. Make sure that the USB cable is not connected.
2. Download the driver from
and follow the instructions
provided with the cable. The installation may take several minutes.
3. When the driver is installed, connect the USB cable to a USB port on your computer.
Windows detects the new device and uses the driver automatically.
4. Verify the COM port for the cable and the status of the cable with
Vaisala USB
Instrument Finder
.
You do not need to uninstall the driver for normal use. To remove the driver files and all Vaisala
USB cable devices, uninstall the entry for
Vaisala USB Instrument Driver
from your computer.
4.2.2 Connecting HMP155 with PuTTY terminal software
You can connect HMP155 with the PuTTY terminal software for Windows (see
) and a USB computer connection cable.
1. Connect the USB serial interface cable to your PC and the RS-485 port of HMP155.
2. Start PuTTY.
Chapter 4 – Operation
37