You can update the parameters with the commands
CRH
,
CT
, and
CTA
. Use
FCRH
only when
you change the sensor.
More information
‣
‣
‣
6.4.1 L command
Use the
L
command to display user adjustment parameters.
Example
>l
Cp offset : 0.0000
Cp gain : 1.0000
T offset : 0.0000
T gain : 1.0000
Ta offset : 0.0000
Ta gain : 1.0000
>
6.4.2 LI command
Use the
LI
command to display and ask for new values for the user adjustment parameters.
If you suspect that something has gone wrong with the calibration and
adjustment, use the
LI
command to return the factory values by entering the
values shown in the following example.
Example
>li
Cp offset : 0.0000 ?
Cp gain : 1.0000 ?
T offset : 0.0000 ?
T gain : 1.0000 ?
Ta offset : 0.0000 ?
Ta gain : 1.0000 ?
>
HMP155 User Guide
M210912EN-D
74