Quantity
Abbreviation
Metric unit
Imperial unit
Temperature
T
°C
°F
Additional T-probe temperature
Ta
°C
°F
The calculated quantities are not available in the passive output version.
Table 34 Quantities calculated from measured quantities
Quantity
Abbreviation
Metric unit
Imperial unit
Dew point / Frost point temperature
(Td/f)
TDF
°C
°F
Dew point temperature (Td)
TD
°C
°F
Mixing ratio (x)
X
g/kg
gr/lb.
Wet bulb temperature (Tw)
TW
°C
°F
7.2 Environmental specifications
Table 35 Operating environment
Property
Description/value
Operating temperature for humidity
measurement
−80 … +60 °C (−112 … +140 °F)
Storage temperature
−80 … +60 °C (−112 … +140 °F)
Operating humidity
0 … 100 %RH
EMC (industrial environment)
EN/IEC 61326-1, EN 55022
Emissions
CISPR 32 Class B (EN 55032)
ESD immunity
IEC 61326-1 (EN 61326-1)
7.3 Accuracy of calculated variables
The accuracy of the calculated variables depend on the calibration accuracy of the humidity
and temperature sensors. Here the accuracies are given for ± 2 %RH and ± 0.2 °C.
HMP155 User Guide
M210912EN-D
82
Summary of Contents for HUMICAP HMP155
Page 1: ...M210912EN D User Guide Vaisala HUMICAP Humidity and Temperature Probe HMP155 ...
Page 8: ...HMP155 User Guide M210912EN D 6 ...
Page 94: ......
Page 98: ...HMP155 User Guide M210912EN D 96 ...
Page 100: ......
Page 101: ......
Page 102: ...www vaisala com ...