5. Maintenance
5.1 HMP155 maintenance
Preventive maintenance for HMP155 includes:
• Checking that the probe is securely mounted.
• Checking that the radiation shield and cable are intact.
• Cleaning the radiation shield using a soft, lint-free cloth moistened with mild detergent.
• Changing the probe filter at least once a year.
If the measurement site is exposed to air pollution and dust, the probe filter
may need to be changed more frequently.
• Sending the sensor to Vaisala for calibration once a year or if the sensor readings seem
inaccurate.
5.1.1 Cleaning
Clean the probe with a soft, lint-free cloth moistened with mild detergent.
5.1.2 Changing HMP155 probe filter
Over time, the probe filter becomes contaminated with matter that you cannot remove by
cleaning. When this happens, the response time becomes longer and measurement accuracy
decreases.
The Pt100 temperature sensor is soldered to the probe. Do not try
to remove it. Due to the operating principles of the sensor, no maintenance is
required.
CAUTION!
1. Remove the probe from inside the radiation shield.
HMP155 User Guide
M210912EN-D
62