Description
Value
At +20 … +60 °C (+68 … +140 °F)
±(0.055 + 0.0057 × |temperature|) °C
Accuracy with passive (resistive) output
According to Tolerance Class AA IEC 60751
±(0.1 + 0.0017 × |temperature|) °C
Accuracy with RS-485 output
At −80 … +20 °C (−112 … +68 °F)
±(0.176 - 0.0028 × |temperature|) °C
At +20 … +60 °C (+68 … +140 °F)
±(0.07 + 0.0025 × |temperature|) °C
1)
Tolerance Class AA IEC 60751 corresponds to IEC 751 1/3 Class B
Figure 13 HMP155 accuracy over temperature range: voltage and RS
‑
485
Table 32 Dew point calculation specifications
Property
Description/value
Accuracy at −20 … +40 °C (−4 … +104 °F)
±0.6 at 90 … 100 %RH
Table 33 Quantities measured by HMP155
Quantity
Abbreviation
Metric unit
Imperial unit
Relative humidity
RH
%RH
%RH
Chapter 7 – Technical data
81
Summary of Contents for HUMICAP HMP155