Command
Description
CTA
One or two-point additional temperature probe calibration and
adjustment.
FCRH
One or two-point relative humidity calibration and adjustment after
sensor change
CTEXT
Type text to calibration information field
L
Display user adjustment parameters
LI
Display new values for user adjustment parameters
Table 19 Analog output commands
Command
Description
AMODE
View analog output modes
ASEL
Set or view analog output quantities and scaling
AERR
Change analog error output values
ATEST
Force analog outputs to entered values
Table 20 Other commands
Command
Description
?
Output information about the device
??
Output information about the device in POLL mode
ERRS
List present probe errors
FILT
[0.1 … 1]
Set the result filtering
HELP
List available commands
PRES
[hPa]
Set the value for pressure compensations
RESET
Reset the probe
VERS
Display software version information
XHEAT
Sensor heating
XPRES
[bar]
Set temporary value for pressure compensations
Chapter 4 – Operation