Modifier
Description
ERR
Error flags for T, Ta, RH, MEM; [0000 ... 1111]; 0 = no error, 1 =error
T = temperature measurement error
Ta = measurement error in the additional T-probe
RH = humidity measurement error
MEM = memory error
For a more detailed error message, use the
ERRS
command
and see
STAT
Probe heating status in 1 character field, for example:
N = no heating
h = warmed probe active
H = purge heating active
S = purge cooling active
X = extra sensor heating active
SNUM
Probe serial number
TIME
Time [hh:mm:ss]
>form "Temperature=" 5.2 t #r#n
OK
>send
Temperature= 24.23
>
>form "Twet=" 6.3 tw U3 #t "T=" t U3 #r#n
OK
>send
Twet= 11.290'C T= 24.231'C
>
>form 5.1 rh #t t #t tdf #r#n
OK
>send
15.6 24.2 -3.1
>
FORM /
command returns the default output format. The default output format depends on
the device configuration.
FORM
command without formatter string returns the current output
format of the probe.
HMP155 User Guide
M210912EN-D
48
Summary of Contents for HUMICAP HMP155
Page 1: ...M210912EN D User Guide Vaisala HUMICAP Humidity and Temperature Probe HMP155 ...
Page 8: ...HMP155 User Guide M210912EN D 6 ...
Page 94: ......
Page 98: ...HMP155 User Guide M210912EN D 96 ...
Page 100: ......
Page 101: ......
Page 102: ...www vaisala com ...