Example
>intv 1 min
Interval : 1 min
>
4.4.4 SEND command
Use the
SEND
ADDR
command, where
ADDR
is the address of the probe, to view the reading
once in the STOP mode.
You can use
SEND
without the address in the STOP mode. The output format
depends on the parameters the probe can print.
The following example shows
SEND
without a parameter.
>send
RH= 24.9 %RH T= 22.1 'C
>
4.4.5 SMODE command
Use the
SMODE
command to set the user port to permanent STOP, RUN, POLL, or SEND mode.
SMODE
[xxxx]
The output mode you select is activated after power outages.
Parameter
Description
xxxx
STOP, RUN, POLL, or SEND
Table 21 Output modes
Mode
Output
Available commands
STOP
Only with the
SEND
command
All (default mode)
RUN
Automatic output
S
POLL
Only with the
SEND
[
addr
] command
??
SEND
[
addr
]
OPEN
[
addr
]
Chapter 4 – Operation