42
20901224_EN_VACUU·PURE 10C_V1.8_221122
Installation and connection
Connect the coolant
A vapor condenser (EK) has a connection for coolant. Water or
the liquid from a chiller, for example, are suitable coolants.
>
The inlet pressure of the coolant at the vapor condenser
must be less than 6 bar (87 psi).
>
A coolant valve may only be installed in the feed line; the
coolant drain must be clear and non‑pressurized.
(2)
(2)
(1)
(1)
IN
IN
OUT
OUT
1.
1.
Fasten both hose shafts
(2)
(2)
to the condenser with union
nuts
(1)
(1) as shown in the diagram.
2.
2.
Attach the hoses for the coolant as shown:
IN = inlet, OUT = outlet.
3.
3.
Secure the hoses,
e.g., with hose clips.
Limitation of use
(US)
Coolant connection max.
pressure
6 bar
87 psi
Example
Coolant connection
on the EK
IMPORTANT!