39
20901224_EN_VACUU·PURE 10C_V1.8_221122
Installation and connection
4.3.2
4.3.2 Outlet connection (OUT)
Outlet connection (OUT)
WARNING
Risk of bursting due to overpressure inside the
outlet line.
Inadmissibly high pressure in the outlet line can cause
the vacuum pump to burst or damage seals.
>
The outlet line (exhaust gas, gas outlet) must always
be clear and non‑pressurized.
>
Always route the outlet line with a fall or take
measures to prevent condensate from flowing back
into the vacuum pump.
>
Observe the maximum admissible pressures and
pressure differences.
WARNING
Danger of bursting due to closed vacuum pump
outlet.
A closed vacuum pump outlet (blind flange) leads to
inadmissibly high pressure at the vacuum pump outlet
and can cause the vacuum pump to burst or damage
seals.
>
Remove the blind flange (transport lock) from the
pump outlet flange before starting the pump.
CAUTION
Excess pressure at the outlet can cause
pumped media to escape.
If the outlet is blocked, pumped media can escape from
the vacuum pump via the sealing gas supply and cause
personal injury and/or pump damage.
>
Do not block the outlet. Do not kink the outlet line.
>
Do not install a shut‑off valve in the outlet line.
>
Use an outlet line with a sufficient cross‑section.
Connect the outlet
line to the outlet