Cleaning and maintenance
3
3.
3. Gradually spray the distilled water directly into the pump
inlet with the drip or spray bottle.
4.
4.
Rinse gradually with about 200 ml of distilled water.
5.
5. If necessary, repeat the cleaning process until all residue is
rinsed out of the vacuum pump.
6.
6. After cleaning is complete, allow the vacuum pump to run
with the inlet open for at least 60 minutes to convey the
remaining water from the vacuum pump.