63
20901224_EN_VACUU·PURE 10C_V1.8_221122
Error messages
Error
Error
`
Possible cause
3
Remedy
Personnel
Warning
flashing rate
5x
`
Supply voltage for
control board in
critical range.
3
Remove any
connected
VACUU·BUS
accessories if too
many or replace if
faulty.
Specialist
Warning
flashing rate
6x
`
Message:
VACUU·BUS
accessories (vacuum
sensor overpressure).
3
Check pressure in
system and reduce
if necessary.
3
Check vacuum
sensor, calibrate if
necessary. Replace
defective sensor.
Operator
`
Message:
VACUU·BUS
accessories (cooling
period for level
sensor active).
3
Cooling period
for level sensor is
running (5 min): No
action necessary.
3
Extend cooling
period for level
sensor.
3
Empty full receiving
flask.
`
Message:
VACUU·BUS
accessories (level
sensor triggers
although level
has not yet been
reached).
3
Calibrate level
sensor or replace
defective level
sensor.
Specialist
Warning
flashing rate
7x
`
Other warnings.
3
Send in vacuum
pump.
Resp.
specialist
Error – Cause –
Remedy
Summary of Contents for VACUU-PURE 10C
Page 1: ...Instructions for use OI no 20901224 Original instructions EN VACUU PURE 10C Vacuum pump ...
Page 12: ...12 20901224_EN_VACUU PURE 10C_V1 8_221122 Introduction ...
Page 93: ...93 20901224_EN_VACUU PURE 10C_V1 8_221122 ...
Page 94: ...94 20901224_EN_VACUU PURE 10C_V1 8_221122 ...